User's Manual

10 • Error Messages Troubleshooting and Diagnostics
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1108 10-8
117 Fluidics Fluid Level Sense
Error.
Droplets, bubbles, or fog
on chamber walls.
Bad fluid level sensor(s).
Bad cable.
Fluid level sensor drive
circuits faulty.
Check the chamber and the
bag for over-inflation.
Select Next Case.
118 Fluidics Irrigation Valve
error.
Valve failure.
Drive circuit failure.
Cycle power.
119 Fluidics Pinch Valve error. Valve failure.
Drive circuit failure.
Cycle power.
120 Fluidics Silicone Extraction
Valve Fault
Valve failure.
Drive Circuit failure.
Disable Venturi or select
Next Case.
121 Fluidics Pack Alignment
Valve Fault.
Pump failure.
Drive circuit failure.
Reinsert the pack and
reprime.
122 Fluidics Dump Valve Fault. Valve failure.
Drive circuit failure.
Disable Venturi or select
Next Case.
123 Fluidics Venturi Valve
Fault.
Valve failure.
Drive circuit failure.
Disconnect compressed air.
Reprime the system.
124 Fluidics Rotary Vane Fault. Pump failure.
Drive circuit failure.
Connect compressed air.
Reprime the system.
125 Fluidics Pack Valve error. Valve failure.
Drive circuit failure.
Cycle power.
127 Fluidics Valve Vent error. Irrigation blocked.
1. End the current Case.
2. Start a New Case.
3. Reprime the System.
4. Turn the system off, wait a
few seconds, then turn the
system on.
5. Install a new tubing
cassette
6. If the error does not clear or
if the error continues to
occur, document the Error
Message and contact AMO
for technical service.
128 Fluidics Pack Loading
error.
Pack not properly seated. Reload the tubing cassette.
Reprime.
129 Fluidics Proportional Valve
Fault.
Valve failure.
Drive circuit failure.
Disconnect compressed air.
Reprime the system.
Error
Number
Message Probable Cause Corrective Action