User's Manual

10 • Error Messages Troubleshooting and Diagnostics
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1108 10-10
148 Pressure to pack capture
mechanism is less than 60
psi.
Leak Reinsert the pack and
reprime.
149 Unloaded strain gauge
reading is too high or too
low.
Pack not loaded correctly. Reinsert the pack and
reprime.
201 Phaco Communication
error.
Bad data sent from host. Retune or cycle power.
202 Phaco power error. Hardware failure (phaco
driver).
Retune or cycle power.
203 Phaco handpiece error. Bad handpiece. Retune or cycle power.
Replace the phaco handpiece.
204 Phaco handpiece error. Wire in handpiece broken. Retune or cycle power.
Replace the phaco handpiece.
207 Incompatible Handpiece
Error.
Incompatible handpiece. 1. Turn off Ellips™
Technology.
2. Attach the Ellips™
handpiece.
3. Tune the handpiece.
210 Phaco RAM error. Chip is bad. Cycle power.
211 Phaco ROM error. Bad 196 processor. Cycle power.
212 Phaco controller timeout. Host stopped writing to
the Phaco controller.
Cycle power.
281 Phaco communication
error.
Software bug or Hardware
failure.
Cycle power.
282 Phaco error. Bad handpiece. Check handpiece and retune.
283 Phaco error. Driver failure. Check handpiece and retune.
Replace the handpiece and
tune.
284 Phaco power supply error. Bad Power Supply. Cycle power or call service.
285 Check for loose handpiece
and retune.
Tip is loose. Check for loose handpiece
tip and retune.
Replace the handpiece and
tune.
286 Phaco handpiece impedance
error.
Bad handpiece. Check handpiece and retune.
Replace the handpiece and
tune.
Error
Number
Message Probable Cause Corrective Action