User's Manual

12 • Specifications
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1108 12-3
Environmental
Specifications
Table 12.2 – Environmental Specifications
Electrical
Specifications
Table 12.3 – Electrical Specifications
Diathermy
Specifications
Table 12.4 – Diathermy Specifications
Irrigation and
Aspiration
Specifications
Table 12.5 – Irrigation and Aspiration Specifications
Environmental
Specifications
Operating Temperature
Humidity
10 to 40°C
Up to 95% RH,
non-condensing
Storage/Transportation
Conditions
Storage Temperature
Humidity
-20 to 60°C
Up to 95% RH,
non-condensing
Voltage Frequency Rated Power Fuse Rating Enclosure
Current Leakage
System
100/120/240
Vac
50/60 Hz 750 VA 6.3A, 250V,
Bussman
GDA
<500 uA
Diathermy
Specifications
Power Adjustment
Diathermy Power (maximum power
into rated load)
Diathermy Type
5 to 100%, in 5% increments
8.5 Watts into 350 ohms 386 KHz
Bipolar
IA Specifications
Peristaltic Pump Vacuum Level 0–650 mmHg in 5 mmHg increments
Pump Flow 0 to 60 cc/minute
Venturi Pump Vacuum Level 0-600 mmHg in 5 mmHg increments