User's Manual

3 • System Setup
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1108 3-17
3. Assemble the phaco handpiece as shown below. You need the handpiece,
titanium phaco tip, the appropriate tip wrench, one of the infusion sleeves and
the test chamber.
CAUTION: NEVER ATTEMPT TO STRAIGHTEN A BENT
NEEDLE. THIS MIGHT PRODUCE A BROKEN TIP WHEN
ULTRASOUND IS APPLIED.
Figure 3.2 – Phaco Handpiece Assembly
4. Attach the connector end of the handpiece to the phaco receptacle on the front
of the WHITESTAR SIGNATURE™ System. Make sure there is no
moisture on the connectors prior to connecting. Moisture prevents the
handpiece from operating properly.
Figure 3.3 – Ellips Handpiece
Note: The Ellips™ handpiece can be used with WHITESTAR
®
Technology and
Ellips™ Technology phaco settings.
1. Test Chamber
2. Infusion Sleeve
3. Handpiece with Tip
12 3