User's Manual

3 • System Setup
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1108 3-25
Programmable IV
Pole
The Programmable IV Pole is controlled by the Up and Down arrows on the upper
right of the touch screen, next to the bottle height indicator. The buttons on the
remote control and the switch on the side of the console can also be used to control
the IV Pole. These controls are used to raise and lower the pole, and the height is
indicated in the Programmable IV Pole screen. The Programmable IV Pole moves
at a rate of approximately 6 cm (2 inches) per second.
The Programmable IV Pole is adjustable from 0 to 104 centimeters, and can be set
for either inches or centimeters. The height measurement is relative to the distance
from the irrigation valve to the center of the drip chamber. The Programmable IV
Pole height for each fluidic mode or submode (PHACO, IA, VIT) is saved in the
WHITESTAR SIGNATURE™ System memory. A Maximum IV Pole height can
be set on the Diagnostics screen.
When a surgery mode is selected, the Programmable Power IV Pole automatically
moves to the preset height. To manually adjust the IV pole height, use the Up and
Down arrows on the touch screen. Manual adjustments to the IV pole can also be
made by pressing the rocker switch located on the side of the console. If a
maximum height has been set, the IV pole will not move above that height.