User's Manual

5 • Anterior Segment Surgery Operating Modes
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1108 5-5
The irrigation tubing runs through the tubing cassette and irrigation is controlled by
the footpedal. When the footpedal is pressed, the pinch valve is opened and the
irrigation fluid flows. Irrigation runs in footpedal Positions 1, 2 and 3.
To program IA submodes:
1. On the left panel, press IA.
2. Select the desired IA submode. IA 1 is selected by default.
3. Press the Up or Down arrows in the control panels on the main panel to
increase or decrease Aspiration Rate or Vacuum.
Note: If the Venturi pump is On, only the Vacuum Settings are shown. To turn
the Venturi pump setting on, use the IA Submode Configuration screen.
4. Press the Settings button in the control panels on the main panel to change
Aspiration Mode or Vacuum Mode,
5. Press Finished to close the Settings window.
6. After all the submodes are programmed, select another mode.
7. If no other changes are needed press End Case to access the SAVE buttons and
save the settings.
Figure 5.4 – Irrigation/Aspiration Mode