User's Manual

5 • Anterior Segment Surgery Operating Modes
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1108 5-26
Table 5.8 – PHACO – AMO Default Settings
Parameter Available
Settings
AMO Default Settings
Phaco 1 Phaco 2 Phaco 3 Phaco 4
Bottle Height 0–104cm
0–42 inches
66 cm
26 inches
76 cm
30 inches
76 cm
30 inches
76 cm
30 inches
Continuous
Irrigation
On, Off Off Off Off Off
Pump Selection Venturi
Peristaltic
Peristaltic Peristaltic Peristaltic Peristaltic
Max Vacuum
Peristaltic
0–650 mmHg 75 mmHg 300 mmHg 400 mmHg 350 mmHg
Max Vacuum
Venturi
0-600 mmHg 75 mmHg 300 mmHg 400 mmHg 350 mmHg
Venturi Pump On, Off Off Off Off Off
Passive Reflux Yes, No
Fusion™
Fluidics Pack
Yes Yes Yes Yes
Fusion™ Dual
Pump Pack
Yes Yes Yes Yes
Initial Mode/
Submode
Any of the
Submode
Names
Phaco 1 Phaco 1 Phaco 1 Phaco 1
Max Flow/
Aspiration Rate,
Unoccluded
(Peristaltic)
10–60 cc/min 18 cc/min 24 cc/min 18 cc/min 24 cc/min
Fluidics Mode Panel, Linear Panel
Aspiration,
Linear
Vacuum
Panel
Aspiration,
Panel
Vacuum
Panel
Aspiration,
Linear
Vacuum
Panel
Aspiration,
Linear
Vacuum
Occlusion
Threshold
0–Max Vac. 25 mmHg 140 mmHg 60 mmHg 80 mmHg
OCCLUSION
MODE
TM
Phaco
On, Off Off On On Off
CASE Mode On, Off Off Off On On
CASE Upper
Threshold.
0–Max Vac. 100 mmHg 250 mmHg 325 mmHg 300 mmHg
CASE Lower
Threshold.
0–Max Vac. 70 mmHg 175 mmHg 200 mmHg 200 mmHg
Up Time 0–2000 ms 300 ms 300 ms 300 ms 300 ms
CASE Vac 120–650
mmHg
85 mmHg 225 mmHg 250 mmHg 250 mmHg
CASE
One-Touch
-2, -1, STD,
+1, +2
STD STD STD STD
Peristaltic Pump
Ramp
10–100% 50% 60% 50% 70%
Power Delivery
Mode
(Unoccluded and
Occluded)
Continuous
Short Pulse
Long Pulse
Low Power
Pulse
High Power
Pulse
Single Burst
Multi-Burst
Continuous
Burst
Continuous Continuous Continuous Continuous
Short Pulse Rate 1–14 pps 6 pps 6 pps 6 pps 6 pps
Long Pulse Rate 1–6 pps 4 pps 4 pps 4 pps 4 pps