User's Manual

5 • Anterior Segment Surgery Operating Modes
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1108 5-27
Table 5.9 – Vitrectomy – AMO Default Settings
WHITESTAR
®
Technology
Duty Cycles
2–30 On Time
2–30 Off Time
Predefined: 6/
12, 4/8, 6/4, 6/
8, 8/4, 4/24, 6/
28, 6/24,
8/24, 6/18
6/12 (33%) 6/12 (33%) 6/12 (33%) 6/12 (33%)
Max Aspiration
Rate Occluded
10–60 cc/min 18 cc/min 28 cc/min 14 cc/min 24 cc/min
Max Phaco
Power Occluded
0–100% 40% 30% 10% 40%
Max Ellips™
Phaco Power
Occluded
On, Off Off Off Off Off
Max Phaco
Power
Unoccluded
0–100% 40% 20% 5% 30%
Max Ellips™
Phaco Power
Unoccluded
On, Off Off Off Off Off
Power Delivery
Type Occluded
Linear, Panel Linear Linear Linear Linear
Customized to Each Surgeon Program/Setup AMO Default Settings
Parameter Available
Settings
VIT 1 VIT 2
Bottle Height 0
104 cm
0
42 inches
30 cm
12 inches
30 cm
12 inches
Continuous Irrigation On, Off Off Off
Pump Selection Venturi
Peristaltic
Peristaltic Peristaltic
Max Vacuum/Mode
(Peristaltic)
0
650 mmHg
Panel, Linear
225 mmHg
Panel
225 mmHg
Panel
Max Aspiration Rate/
Mode
0
60 cc/min
Panel, Linear
18 cc/min
Panel
12 cc/min
Panel
Fluidic Mode Panel, Linear Panel
Aspiration
Panel
Vacuum
Panel
Aspiration
Panel
Vacuum
Cut Rate 50
2500 cpm 250 cpm 450 cpm
Footpedal Mode ICA, IAC ICA IAC
Pump Ramp
Threshold
10 - 100% 100% 100%
Venturi Pump On, Off Off Off
Parameter Available
Settings
AMO Default Settings