User's Manual

5 • Anterior Segment Surgery Operating Modes
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1008 5-24
Table 5.5 – Footpedal Threshold AMO Default Settings
*Switch Assignments
Table 5.6 – Diathermy – AMO Default Settings
Footpedal
Type
Parameter Available
Settings
AMO Default
Settings
Single Linear Pitch
Regions
1–100% P1=5%
P2=30%
P3=60%
Switch
Position
Switch1= *
Switch 2= *
Switch1=Off
Switch 2=Off
Right= *
Left= *
Right=Off
Left=Off
Feedback On, Off Off
Advanced
Control Pedal
(Dual Linear)
Primary Yaw * Reflux
Pitch Region 1-100% P1=5%
P2=30%
P3=60%
Yaw Position 5-100% Left=50%
Right=50%
Switch
Position
Switch1= *
Switch 2= *
Switch1=Off
Switch 2=Off
Yaw
Threshold
5% - 100% 5%
Feedback On, Off Off
Surgeon Program Up Surgeon Program Down Previous Major Mode Next Major Mode Previous Active Mode
Next Active Mode Previous Sub Mode Next Sub Mode Toggle SMC Record Bottle Up
Bottle Down Reflux Continuous Irrigation 1-Touch Up (Phaco
Only)
1-Touch Down (Phaco
Only)
Toggle Case (Phaco
Only)
Single Vit Cut
(Vitrectomy Only)
Parameter Available
Settings
AMO Default Settings
DIA 1 DIA 2
Max Power 5–100% 30% 30%
Power Delivery
Type
Linear, Burst,
Panel
Linear Burst
Initial Mode/
Submode
Any of the
Submode
Names
Phaco 1 Phaco 1