User's Manual

5 • Anterior Segment Surgery Operating Modes
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1008 5-25
Table 5.7 – I/A – AMO Default Settings
Parameter Available
Settings
AMO Default Settings
IA1 IA2 IA3
Continuous
Irrigation
On, Off Off Off Off
Pump
Selection
Venturi
Peristaltic
Peristaltic Peristaltic Peristaltic
Bottle Height 0–104 cm
0–42 inches
76 cm
30 inches
50 cm
20 inches
50 cm
20 inches
Max Vacuum 0–650
mmHg
500
mmHg
500
mmHg
15 mmHg
Max
Aspiration/
Mode
(Peristaltic)
0–60 cc/min 26 cc/min
Panel
26 cc/min
Linear
6 cc/min
Panel
Peristaltic
Pump Ramp
Threshold
10–100% 80% 80% 80%
Venturi Pump On, Off Off Off Off
Passive Reflux
(Fusion™
Fluidics Pack)
Yes, No Yes Yes Yes
Passive Reflux
(Fusion™
Dual Pump
Pack)
Yes, No Yes Yes Yes
Initial Mode/
Submode
Any of the
Submode
Names
Phaco 1 Phaco 1 Phaco 1
Fluidic Mode Panel, Linear Panel
Aspiration
Linear
Vacuum
Linear
Aspiration
Panel
Vacuum
Panel
Aspiration
Linear
Vacuum