User's Manual

RX Only NGP Z370500 Rev. B 1116
6-11
WHITESTAR SIGNATURE PRO System
6 System Configuration
Touch Screen
Calibration
You must calibrate the system touch screen as part of the system setup. Follow
these steps to calibrate the touch screen.
1. On the main screen, press Configuration.
2. Press Touch Screen Calibration.
Figure 6.10 Touch Screen Target Circle
3. Press the arrow on the screen. When released, the arrow moves to the next
calibration point.
4. Repeat for all the calibration points.
5. Press Confirm at the completion of the calibration process.