User's Manual

WHITESTAR SIGNATURE PRO System
8 Error Messages Troubleshooting and Diagnostics
RX Only NGP Z370500 Rev. B 1116
8-2
Most Common
User-Correctable
Problems
Use the information in this section if you are not successful with the system
check-out. Consult this section to resolve the problem before you call AMO for
technical service:
Make sure that you plugged the system in to a power receptacle.
Make sure that there is electrical power to the receptacle.
If there is no
phacoemulsification,
make sure that the phaco needle is tight on the
handpiece.
If there is no
phacoemulsification,
make sure that the phaco needle is compatible
with the handpiece (for example, non-AMO phaco needle on an AMO
handpiece).
If there is no phacoemulsification, confirm there is no damage to the phaco
needle/handpiece by dropping or misuse.
If no irrigation occurs, shake the drip chamber to confirm that the ball or the
valve moves freely. If there is no rattle sound, replace the drip chamber with
another disposable pack.
Fuse Replacement
Procedure
If the system does not turn on when you press the power switch, you have
confirmed that you attached the power cord properly, and you plugged the cord in,
check to see if the fuse is bad.
Note: To prevent the risk of fire or damage to the instrument, replace the fuses
with the exact type and rating as indicated below (check the voltage sticker
on the back panel of the system to confirm your system voltage):
Table 8.1 – Fuse Specifications
Voltage
Quantity
Fuse
Specifications
Console
100/120/240
2
6.3A, 250V,
Bussman GDA