8 Error Messages Troubleshooting and Diagnostics
RX Only NGP Z370500 Rev. B 1116
8-8
Move the first remote out of range from the console.
Wait for the first remote to go into “sleep” mode.
Pair the new remote.
6. Make sure that you fully charge the batteries before you pair the remote with
the console. Low batteries can cause pairing failures.
7. Charge the batteries if pairing has failed after several attempts.
8. When the batteries are charging:
The BLUETOOTH is off.
You cannot use the remote.
You cannot pair the remote.
The foot pedal is not operating properly
1. Go to the Configuration screen and perform a Foot Pedal Test.
2. Confirm that you attached the foot pedal cable at the back of the console.
3. Perform a Foot Pedal Calibration, for either the Advanced Control Pedal or
the Single Linear foot pedal.
The IV Pole does not respond
1. The IV pole might have reached the maximum or the minimum height.
2. Attempt a programmable IV pole height adjustment with the touch screen,
remote, or the rocker switch on the side of the system.
The Touch Screen Does Not Respond
Perform the Touch Screen Calibration procedure as described in “Touch Screen
Calibration” on page 6-11.
Priming Errors
1. Check the pack loading, including reloading the pack.
2. Verify that there are no kinks, clogs, or loose fittings.
3. Replace the handpiece and the tip and prime.
4. Replace the pack.
5. Check the test chamber for proper installation and any leaks.
6. Contact AMO Technical Service to check the vacuum.