User's Manual

WHITESTAR SIGNATURE PRO System
8 Error Messages Troubleshooting and Diagnostics
RX Only NGP Z370500 Rev. B 1116
8-10
Using large amounts of fluid
1. Check the incision size.
2. Check the bottle height.
3. Check the flow rate (pump speed too high).
4. Check that no fluid enters the collection bag when you do not use irrigation.
5. Reseat or replace the tubing.
Aspiration No aspiration
1. Make sure you have the appropriate mode selected on the screen.
2. Check for kinks or clogs in the tubing.
3. Check the tubing connection to the handpiece.
4. Make sure there are no obstructions in the handpiece.
5. Press the foot pedal to position 2 and check the pump function.
Poor Aspiration
1. Check the flow rate.
2. Check the foot pedal operation.
3. Check for kinks or clogs in the tubing.
4. Make sure there are no obstructions in the handpiece.
5. Check the tubing connection to the handpiece.
6. Check the I/A handpiece o-rings for excessive wear. Replace the o-rings, if
needed.