The system shows status, warning, and error messages on the monitor.
Alerts (Gray) show at the top of the screen. You do not need to clear an alert as with
an error message. An alert, for example, can be: Not Primed/Not Tuned.
Error messages show at the top of the screen. Press button to open the
corrective action for that error. After you correct the error, press to clear the
error message from the screen.
The message can show possible solutions to the error or recommendations to clear
the error. The messages can indicate the available options if a component or
subsystem fails.
The list of messages in the following pages have corrective actions that you can
take to clear the error.
Error Messages
101 Fluidics communication error.
Probable cause: Invalid data or communication error.
Corrective action:
1. Select End Case.
2. Select Next Case.
3. If the error does not clear select End Case.
4. Select Shutdown.
5. Select Yes and wait for the system to power off.
6. Start up the system.
7. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.
102 Fluidics write error.
Probable cause: The Microcontroller, SPI, SPI Bus, EEPROM, or ADC is bad.
Corrective action:
1. Select End Case.
2. Select Shutdown.
3. Select Yes and wait for the system to power off.
4. Start up the system.
5. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.