8 Error Messages Troubleshooting and Diagnostics
RX Only NGP Z370500 Rev. B 1116
8-16
112 Fluidics Master communication error
Probable cause: There is no communication between the Fluidics Controller and
the Instrument Host.
Corrective action:
1. Select End Case.
2. Select Shutdown.
3. Select Yes and wait for the system to power off.
4. Start up the system.
5. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.
114 Fluidics DAC fault.
Probable cause: Circuit failure of the I2C bus, DAC, or micro.
Corrective action:
1. Select End Case.
2. Select Shutdown.
3. Select Yes and wait for the system to power off.
4. Start up the system.
5. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.
118 Fluidics irrigation valve error.
Probable cause: The valve or drive circuit has failed.
Corrective action:
1. Select End Case.
2. Select Shutdown.
3. Select Yes and wait for the system to power off.
4. Start up the system.
5. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.