8 Error Messages Troubleshooting and Diagnostics
RX Only NGP Z370500 Rev. B 1116
8-18
125 Fluidics pack valve error.
Probable cause: Valve or drive circuit failure.
Corrective action:
1. Select End Case.
2. Select Shutdown.
3. Select Yes and wait for the system to power off.
4. Start up the system.
5. Eject the
FUSION
Fluidics pack.
6. Insert the
FUSION
Fluidics pack.
7. Insert a new
FUSION
Fluidics pack.
8. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.
126 Fluidics motor vent error.
Probable cause: Vacuum chamber is blocked.
Corrective action:
1. Select End Case.
2. Select Next Case.
3. Replace the FUSION Fluidics pack.
4. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.
129 Fluidics proportional valve fault.
Probable cause: Valve or drive circuit failure.
Corrective action:
1. Disconnect the external air supply.
2. Prime the system again.
3. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.