8 Error Messages Troubleshooting and Diagnostics
RX Only NGP Z370500 Rev. B 1116
8-19
130 Fluidics encoder error.
Probable cause:
There is a tubing pack loading problem.
Encoder, decoder or stepper driver circuit failed.
Corrective action:
1. Select End Case.
2. Select New Case.
3. If the error does not clear select End Case.
4. Select Shutdown.
5. Select Yes and wait for the system to power off.
6. Start up the system.
7. Remove the tubing pack.
8. Reinstall the tubing pack.
9. Install a new tubing pack.
10. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.
131 Fluidics mode error
Probable cause:
Instrument Host malfunction
Bad SSC Board
Bad Foot pedal
Corrective action:
1. Select End Case.
2. Select Shutdown.
3. Select Yes and wait for the system to power off.
4. Start up the system.
5. Disconnect the foot pedal from the system.
6. Connect the foot pedal to the system.
7. Press and release the foot pedal at least three times.
8. Run the Self Test.
9. Replace the foot pedal.
10. If the error is not corrected or if the error continues to occur, document the error
message and contact AMO for technical service.