211 Phaco ROM error.
Probable cause: Bad Microcontroller.
Corrective action:
1. Select End Case.
2. Select Shutdown.
3. Select Yes and wait for the system to power off.
4. Start up the system.
5. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.
212 Phaco controller timeout.
Probable cause: Instrument Host failure or bad Microcontroller.
Corrective action:
1. Select End Case.
2. Select Shutdown.
3. Select Yes and wait for the system to power off.
4. Start up the system.
5. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.
213 SPI error.
Probable cause: SPI Microcontroller failure.
Corrective action:
1. Select End Case.
2. Select Shutdown.
3. Select Yes and wait for the system to power off.
4. Start up the system.
5. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.