214 SPI error.
Probable cause: SPI Microcontroller failure.
Corrective action:
1. Select End Case.
2. Select Shutdown.
3. Select Yes and wait for the system to power off.
4. Start up the system.
5. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.
281 Phaco communication error.
Probable cause: Software or hardware failure.
Corrective action:
1. Select End Case.
2. Select Shutdown.
3. Select Yes and wait for the system to power off.
4. Start up the system.
5. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.
283 Phaco error.
Probable cause: Phaco Drive failure.
Corrective action:
1. Check the handpiece connections.
2. Check the handpiece. Replace the tip if needed.
3. Retune the handpiece.
4. If the error still does not clear, replace the handpiece with a new handpiece, and
tune the handpiece.
5. If the error still does not clear or if the error continues to occur, document the
error message and contact AMO for technical service.