287 Diathermy error.
Probable cause: Diathermy Driver failure.
Corrective action:
1. Check the diathermy cord connections.
2. Select End Case.
3. Select Shutdown.
4. Select Yes and wait for the system to power off.
5. Start up the system.
6. If the error still does not clear or if the error continues to occur, document the
error message and contact AMO for technical service.
290 Foot pedal error.
Probable cause:
The foot pedal has failed.
The user pressed the foot pedal.
Corrective action:
1. Check the foot pedal. Make sure the foot pedal is connected.
2. Select End Case.
3. Select Shutdown.
4. Select Yes and wait for the system to power off.
5. Start up the system.
6. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.
291 Phaco Diathermy power supply error.
Probable cause: Diathermy regulator or driver failure.
Corrective action:
1. Select End Case.
2. Select Shutdown.
3. Select Yes and wait for the system to power off.
4. Start up the system.
5. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.