8 Error Messages Troubleshooting and Diagnostics
RX Only NGP Z370500 Rev. B 1116
8-32
328 Pneumatics system pressure too high.
Probable cause:
The system pressure is too high.
The external air valve does not work.
The piston pump does not work.
The selector valve does not work.
Corrective action:
1. Select End Case.
2. Select Next Case.
3. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.
352 IH Failure.
Probable cause: Hardware not functioning correctly.
Corrective action:
1. Select End Case.
2. Select Shutdown.
3. Select Yes and wait for the system to power off.
4. Start up the system.
5. If the error still does not clear or if the error continues to occur, document the
error message and contact AMO for technical service.
353 IH to GUI communication error.
Probable cause: Software or hardware error.
Corrective action:
1. Select End Case.
2. Select Next Case.
3. If the error still does not clear or if the error continues to occur, document the
error message and contact AMO for technical service.