8 Error Messages Troubleshooting and Diagnostics
RX Only NGP Z370500 Rev. B 1116
8-34
370 IH Phaco read error.
Probable cause: Communication error.
Corrective action:
1. Select End Case.
2. Select Shutdown.
3. Select Yes and wait for the system to power off.
4. Start up the system.
5. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.
371 IH Phaco write error.
Probable cause: Communication error.
Corrective action:
1. Select End Case.
2. Select Shutdown.
3. Select Yes and wait for the system to power off.
4. Start up the system.
5. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.
374 Handpiece removed after being tuned.
Probable cause: Phaco handpiece removed.
Corrective action:
1. Attach the handpiece.
2. Tune the handpiece and continue surgery.
3. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.