User's Manual

WHITESTAR SIGNATURE PRO System
8 Error Messages Troubleshooting and Diagnostics
RX Only NGP Z370500 Rev. B 1116
8-37
503 Prime low vacuum error.
Probable cause: Hardware failure.
Corrective action:
1. Remove the tubing pack.
2. Reinstall the tubing pack and reprime.
3. If the error does not clear, replace the tubing pack with a new tubing pack and
prime the system.
4. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.
506 Pack detect failure.
Probable cause: A pack switch failure.
Corrective action:
1. Remove the tubing pack.
2. Reinstall the tubing pack and reprime.
3. If the error does not clear, replace the tubing pack with a new tubing pack and
prime the system.
4. If the error does not clear or if the error continues to occur, document the error
message and contact AMO for technical service.
512 IV Pole communications error.
Probable cause: Communication error.
Corrective action:
1. Prime the system.
2. If the error still does not clear or if the error continues to occur, document the
error message and contact AMO for technical service.