User's Manual

WHITESTAR SIGNATURE PRO System
5 • Program Settings
RX Only NGP Z370500 Rev. B 1116
5-3
Figure 5.2 – Delete Program Confirmation Dialog Box
3. Press Yes.
Edit a Program
Name
1.
Select a program name on the list.
2.
Press .
3.
4.
Enter a different program name or select a new name from the Standard
Names list.
Press Standard Names.
5.
Select a program from the list.
6.
Press Enter.
Program Settings Program settings are used to define your specific settings on the WHITESTAR
SIGNATURE PRO system. There are four major surgical modes (Phaco, I/A,
Vitrectomy (VIT), and Diathermy (DIA)).
For each major mode there are multiple submodes available that you can customize
for each step in your procedure. You can program all or some of the submodes
within each of the major modes. You can use the submode names that come with
the system (Standard Names) or create a specific name.
The primary settings in the submodes are:
Aspiration
Vacuum
Power
Cut Rate (Vitrectomy)
Bottle Height
Pump Type (Peristaltic or Venturi)
Delivery Mode (Panel or Linear)
Linear - Provides a gradual increase based on the foot pedal position.
Panel - Provides a full and immediate increase when the is pressed.