User's Manual

WHITESTAR SIGNATURE PRO System
5 • Program Settings
RX Only NGP Z370500 Rev. B 1116
5-9
Parameter
Available
Settings
Default Settings
Max power
Unoccluded (%)
40
20
5
30
Power delivery
Continuous
Continuous
Continuous
Continuous
WHITESTAR setting
(duty cycle)
On 6/12 (33%)
On 6/12 (33%)
On 6/12 (33%)
On 6/12 (33%)
Occluded mode (%)
40
30
10
40
Power delivery
Continuous
Continuous
Continuous
Continuous
WHITESTAR setting
(duty cycle)
On 6/12 (33%)
On 6/12 (33%)
On 6/12 (33%)
On 6/12 (33%)
Power type, occluded and
unoccluded
Linear
Linear
Linear
Linear
Short pulse rate, occluded
and unoccluded
1–14 pps
6 pps
6 pps
6 pps
6 pps
Long pulse rate, occluded
and unoccluded
1–6 pps
4 pps
4 pps
4 pps
4 pps
Pulse Shape Setting
(WHITESTAR ICE)
On, Off
Off
Off
Off
Off
Pulse Shape Parameters
% Kick Low
5
5
5
5
% Kick High
5
5
5
5
Low Power Limit
0
0
0
0
High Power Limit
80
80
80
80
Foot Pedal Feedback
On, Off
Off
Off Off Off
Initial Mode
On, Off
Phaco (first enabled
submode)
Passive Reflux On, Off On
On On On
Cup Fill 30, 60, 90 cc 30 cc