User's Manual

WHITESTAR
SIGNATURE
PRO
System
4
Surgeon
Settings
RX Only NGP Z370500 Rev. B 1116
4-31
Follow these steps to set the foot pedal thresholds.
Note: In addition to the procedure below, you can access the Advanced Linear
Pedal Summary screen by pressing the foot pedal symbol near the top of
the screen.
1. On the Surgeons and Programs screen, press .
2. Press the Thresholds button under Foot Pedal. The system displays the foot
pedal thresholds screen.
3. Use the up and down arrows to change the threshold for each of the foot pedal
pitch zones. P1 is the first zone, P2 is the second, and P3 is the final
zone.
4. Select Feedback to enable foot pedal feedback. Feedback is a physical and
audible click as you move the foot pedal to the next or previous position. The
foot pedal is now ready for calibration.
5. Press Save to retain the settings.
6. Press Exit to close the Foot Pedal Set Up screen.