User's Manual

9 Clinical Considerations
Effects of Electrosurgical Equipment
9.2 RECOVER BP IABP/VAD System – Operator’s Manual
Effects of Electrosurgical Equipment
The RECOVER BP Console may exhibit the following effects during
the use of electrosurgical equipment:
Display becomes blank or distorted
Alarm sounds
Pumping stops or becomes erratic
The display and the pumping function will resume normal operation
within 10 seconds after the completion of electrosurgery.
Effects of Defibrillation Equipment
WARNING: Be extremely careful when a defibrillator is used on a
patient. Dangerous high voltage is present during defibrillation. Do
NOT touch the Console, patient, table, accessories, cables, or any
connected equipment.
The Console is classified as Type CF defibrillation-proof
(IEC 60601-1) equipment.
The following effects may occur during defibrillation:
Display becomes blank or distorted
Alarm sounds
Pumping stops or becomes erratic
The display and the pumping function will resume normal operation
within 10 seconds after the completion of defibrillation.