10 Installation and Maintenance
Cleaning
10.4 RECOVER BP IABP/VAD System – Operator’s Manual
Cleaning
CAUTION: Do NOT pour liquid, including cleaning solution,
directly on any part of the Console. Doing so can cause electrical
malfunction.
If liquid is accidentally spilled on the Console, be sure to thoroughly
dry the affected area. Wait at least 15 minutes, after drying, before
turning the Console ON. Verify that the Self-Test runs and indicates
that the unit is operating properly.
Console
It is recommended that the Console be cleaned after each use. Turn
the Console OFF before you clean it.
To clean the display/control panel area, use cotton and 70% isopropyl
alcohol, or a soft cloth with soap and water.
To clean the housing, use a damp cloth and a mild detergent solution.
Patient ECG Cable
Clean and disinfect the patient ECG cable after each use by
performing the procedure described in the Instructions for Use
supplied with the cable or by following hospital operating procedures.