User's Manual

System Specifications 12
Electromagnetic Compatibility
RECOVER BP IABP/VAD System – Operator’s Manual 12.11
TABLE 206
Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and the RECOVER BP Console Equipment and Systems that are NOT
Life-
Supporting
The RECOVER BP Console is intended for use in the electromagnetic environment in which
radiated disturbances are controlled. The customer or user of the RECOVER BP Console can
help prevent electromagnetic interference by maintaining a minimum distance between Portable
and Mobile RF Communications Equipment and the RECOVER BP Console as recommended
below, according to the maximum output power of the communications equipment.
Recommended Separation Distances for the RECOVER BP Console
Maximum Output
Power (Watts)
150 KHz to
80 MHz
Pd 1667.1=
80 to 800 MHz
Pd 1667.1=
800 MHz to 2.5 GHz
Pd 3333.2=
0.01
0.11667
0.11667
0.23333
0.1
0.36894
0.36894
0.73785
1
1.1667
1.1667
2.3333
10
3.6894
3.6894
7.3785
100
11.667
11.667
23.333