User's Manual

IABP Support 6
Using the Interface
RECOVER BP IABP/VAD System – Operator’s Manual 6.3
Table 10 IABP Keypad Functions
Feature Use
Control Keys
ON Purges the pneumatic system and starts pumping.
STANDBY
Pauses pumping but keeps the Console ready to immediately resume
pumping. Pumping resumes when ON is pressed.
OFF Stops pumping when pressed twice within 13 seconds.
AUTO
Automatically selects optimal ECG source, ECG gain, AP source,
trigger source, inflation timing, and deflation timing.
MANUAL
Allows you to select ECG source, AP source, trigger source, inflation
timing, and deflation timing.
ECG Selects ECG signal as the trigger source.
PRESSURE Selects AP signal as the trigger source.
OTHER Allows you to select Apace, Vpace, or Internal as the trigger source.
FREEZE Freezes or unfreezes the display.
INFLATE
In MANUAL mode, the left and right arrows adjust the inflation
interval. Inactive in AUTO mode.
DEFLATE
In MANUAL mode, the left and right arrows adjust the deflation
interval. Inactive in AUTO mode.
REFERENCE Up and down arrows select the reference line.
AUG ALARM Up and down arrows set the augmentation alarm level.
AUG Up and down arrows set the augmentation level.
FREQ Selects assist frequency (1:1, 1:2, or 1:3).
INFLATION MARKER Turns inflation marker on or off.
Indicator Lights
ON Green light when ON is pressed.
STANDBY Amber light when STANDBY is pressed.
OFF
Red light flashes when OFF is pressed once (pumping does not
stop). Light stays red when OFF is pressed twice within 13 seconds
(pumping stops).
AUTO Green light when AUTO is pressed.
MANUAL Amber light when MANUAL is pressed.
ECG Green light when ECG is selected as the trigger source.
PRESSURE Green light when AP is selected as the trigger source.
OTHER
Green light when Apace, Vpace, or Internal is selected as the trigger
source.
AUGMENTATION Green light bar indicates augmentation level.
FREQ Green light indicates the selected assist frequency (1:1, 1:2, or 1:3).
****[INTRA-AORTIC BALLOON PUMP label]