IABP Support 6
Initiating Support
RECOVER BP IABP/VAD System – Operator’s Manual 6.11
To Initiate Support in Manual Mode:
WARNING: Due to the potential for thrombus formation, an IAB
must NOT remain dormant.
1 Select “Manual” on the IABP keypad.
2 Refer to the following text to choose the desired option for each
of the following settings:
ECG source and ECG gain
AP source
Trigger source
Inflation and deflation timing
Setting the ECG Source and ECG Gain
WARNING: Do NOT allow the conductive parts of electrodes and
associated connectors to contact any conductive parts and/or earth
ground.
Acquiring a high-quality electrocardiograph (ECG) waveform is an
important factor in achieving optimal triggering. An ECG waveform
can be obtained using either skin electrodes or the output of an
external monitor.
Using Skin Electrodes
1 Using the supplied patient ECG cable, connect each patient lead
wire to a skin electrode. Use only high-quality electrodes.