User's Manual

IABP Support 6
Initiating Support
RECOVER BP IABP/VAD System – Operator’s Manual 6.15
Using an External Monitor
1 Connect the AP high-level input interface cable between the
external patient monitor and the high-level AP input on the side
panel. Make sure that ****[X] is selected on the display.
2 Press the
MENU key and select AP Source > External (see
Figure 23). The left side of the display shows the selected AP
source.
Figure 23 Selecting the AP Source for an External Monitor
3 Check that the arterial pressure waveform is properly displayed.
Selecting the Trigger Source
ECG
Using this setting, the Console triggers on the QRS and ignores
pacemaker pulses. Select this option under TRIGGER SOURCE on
the keypad.