User's Manual
IABP Support 6
IABP Alarms
RECOVER BP IABP/VAD System – Operator’s Manual 6.23
Table 14 IABP White Alarm Messages
WHITE ALARMS Category: Advisory
Alarm Message
Possible Cause
Action
LEAD OFF
ECG Electrode
• ECG lead off.
1. Check ECG leads.
2. Limit movement of
ECG leads.
HELIUM SUPPLY
Low
• Low helium cylinder
pressure.
Replace helium
cylinder.
SYSTEM
Chart Recorder
• Out of paper.
• Door open.
1. Check paper.
2. Close door.
3. Call for service.