RECOVER BP VAD/IABP System – Operator’s Manual 7.5
Recommended Cannulation Method
Refer to the Cannula Instructions for Use (0506-9110).
Pump Type Verification
Be sure that “Ventricle” is displayed as the pump type.
Initiating Support
1 Initiate support as described in the AB5000
Ventricle Instructions
for Use and the AB5000 Ventricle Training Guide.
2 Begin pumping by pressing the
ON key for the appropriate side
(Left or Right). If both sides are being supported, press the Left
ON key first.
Optimizing AB5000 Ventricle Filling
When the Console is operating in normal (full flow) mode, flow rate
should be greater than 2 L/min and typically ranges from 3 to
6 L/min.
To optimize blood flow, be sure that the patient is appropriately
hydrated, with filling pressures within normal ranges. If the patient is
hypovolemic, administer fluids according to hospital protocol.
The Ventricle is filled via vacuum assist. However, avoid raising the
Ventricle above the heart to view it, because this could result in lower
flow.