Reference Manual

BiWaze Cough - User Reference Manual
34
6.6.1 Main Therapy settings
From the Main Therapy screen, the following settings may appear on-screen,
depending on how the device is configured.
Note
: When the device is completely locked the users can only select available
profiles (presets) enabled by HCP, but not adjust any of the settings.
Note
: HCP may lock the device with some adjustment allowed by end user, the
range of settings that can be changed in that mode is small (<5cm and ± 2 secs)
and depends on configuration done by HCP
Setting
Description
Modes and Profiles
Allows you to quickly select a group of predefined
prescription settings under each mode (Manual or Auto).
Oscillations Toggle
Allows you to Enable/Disable the oscillations. Oscillation
creates the pressure pulses delivered to the patient
based on Frequency and Amplitude settings. (see
advanced therapy settings)
Inspiratory Pressure
Allows you to set the Inspiratory Pressure setting from 0
to 70 cmH2O in increments of 1. The Inhale Pressure is
the pressure the patient receives while in the Inhale
phase. User can adjust flow pattern for inspiratory phase
from Advance Therapy Settings.
Inhale Time
Allows you to set the Inhale Time from 0.0 to 5.0
seconds in increments of 0.5. Inhale Time indicates how
long the patient spends in the Inhale phase when in Auto
mode. This setting is not available when Therapy Mode
is set to Manual.
Exhale Pressure
Allows you to set the Exhale Pressure from 0 to -70
cmH2O in increments of 1. Exhale pressure is the
pressure the patient receives while in the Exhale Phase.
Exhale Time
Allows you to set the Exhale Time from 0.0 to 5.0
seconds in increments of 0.5. Exhale Time indicates how
long the patient spends in the Exhale Phase when in
Auto mode. This setting is not available when the mode
is set to Manual.
Pause Pressure
Allows you to set the Pause Pressure from 0 to 30
cmH2O in increments of 1. Exhale pressure is the