Reference Manual

BiWaze Cough - User Reference Manual
36
1,2,3 Increment, decrement, reset the pressure setting
4,5 Increment, decrement time setting
6 Confirm the change
Note
: the pressure and time settings on main screen get auto saved for next
reboot if the therapy was delivered with those settings.
6.6.2 Advanced Therapy settings
Following therapy settings are available when the user touches the Advanced
Settings icon at the bottom left corner of the main screen.
Setting
Description
Add/Remove
Profiles
Allows you to save currently selected settings as profile
as presets for quick select. User can also remove
currently selected profile from this screen.
Inspiratory Trigger
toggle
Allows you to Enable/Disable the inspiratory trigger. If
the inspiratory trigger is on, the pause time setting is
over ridden as the device trigger inspiratory phase from
pause phase of breath only when it detects an inspiratory
effort.
Inspiratory Trigger
Sensitivity
Allows you to choose between sensitivity in range of 1-
10 with 10 being most sensitive. User can update this
setting to adjust sensitivity of trigger.
Inspiratory Flow
This setting allows user to select between low, medium
and high inspiratory flow. For patient comfort its
recommended to use “Low” flow setting for inspiratory
phase of breath.
Oscillation
Amplitude
Allows user to set the oscillation amplitude in range from
1 to 5 with 5 being higher amplitude. Refer to Section:
Oscillation Feature for details
Oscillation
Frequency
Allows user to select frequency of oscillation in range of
5-20Hz with increments of 1