WARNING:
Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the BiWaze Cough System, including cables
specified by the manufacturer. Otherwise, degradation of the performance
of this equipment could result.”
The BiWaze Cough System should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the BiWaze Cough
System should be observed to verify normal operation. If operation is not
normal, the BiWaze Cough System or the other equipment should be
moved.
Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
13.1
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specified below.
The user of this device should make sure it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment -
Guidance
RF emissions CISPR 11
Group 1
The device uses RF energy only for its internal