Reference Manual

BiWaze Cough - User Reference Manual
68
Move the device away from
any sources of heat or hot
ambient temperatures.
Switch off the device and
restart after 15 minutes of
cool down.
If problem persists, call your
health care provider for
servicing the device to
replace the filters and
functional check.
Device does not power up
Battery may be too low.
Connect the device to mains
power and check if the
device powers up.
If problem persists, call your
health care provider for
service.
Technical errors with an
error code number on a Red
strip on the LCD Screen and
device shuts down after few
seconds
Technical error related to
temperatures or other high
priority fault
Try rebooting the device and
if problem persists, call your
health care provider for
service.
Technical errors with an
error code number on a Red
strip on the LCD Screen and
device does not shut down .
User cannot start the
therapy.
Technical error related to
subsystem malfunction
Try rebooting the device and
if problem persists, call your
health care provider for
service.
Information with a self-
explanatory message on the
Informational messages
User can acknowledge and
continue with therapy. Take