Technical data
Farbe: P 021 – Prüfmittelnummer: 201
Battery life   approx. 1000 tests or approx. 1 year (less in 
Acoustic Mode owing to greater power 
requirement)
Interface  Infrared port
Safety class  III
LED/IRED  Class 1
Electromagnetic   This meter meets the electromagnetic 
compatibility  immunity requirements as per ISO 15197 
Annex A. The chosen basis for electrostatic 
discharge immunity testing was basic stan-
dard IEC 61000-4-2 . 
  In addition it meets the electromagnetic 
emissions requirements as per EN 61326. 
Its electromagnetic emission is thus low. 
Interference from other electrically driven 