This message/
symbol appears ...
and means:
You applied the blood drop or control solution
drop too soon.
Turn the meter off and on again. Repeat the
test. Wait until the flashing hand symbol and
blood drop symbol appear in the display and
the meter beeps (provided the beep tone is
turned on) before applying blood or control
solution.
The optical system is dirty.
Clean the optical system (see Chapter 7
“Cleaning the Accu-Chek Integra meter”)
and repeat the test.
The meter was turned off and on again. The
strip is still in the meter.
Turn the meter off. Hold the meter upright
and let the strip drop out, or remove the strip
manually, even if it takes a small amount of
force.
You forcibly removed the test strip during
testing.
Turn the meter off and on again. Repeat the
test with a new strip, this time ejecting it by
means of the ON/OFF button.
This message/
symbol appears ...
and means:
The test strip that has been advanced is
defective.
Hold the meter with the strip pointing down
-
wards. Turn the meter off and let the strip
drop out. Turn the meter on again. Repeat
the test with a new test strip.
The two arrows do not start to flash. The meter
turns off after 5 minutes and the computer dis-
plays a download-related error message.
An error occurred while data was being
downloaded.
Download the data again.
A bright light source is too close to one of the
infrared windows.
Increase the distance to the light source and
re-start the data download.
9.2 Troubleshooting