User`s manual
94
Farbe: PMS 583 – Prüfmittelnummer: 208
95
This message/
symbol appears ...
and means:
  The meter and the infrared cable are too 
far away from one another or not correctly 
aligned.
  Decrease the distance between the infrared 
windows of the two devices to less than 
10 cm. 
  Make sure the two infrared windows are 
pointing towards one another. Re-start the 
data download.
  A computer software error occurred. 
Read the Manual for the download software.
  Electrostatic discharge 
Turn the meter off and on again.
  Meter is defective. 
Please contact Customer Support (see Chap
-
ter 19 for details).
10  Technical data
Meter type
Accu-Chek Integra
Catalogue number/serial 
number
See type plate on the back of the 
meter.
Test principle
Determination of glucose in fresh 
capillary blood by reflectance pho
-
tometry. When using different speci
-
men material please refer to the 
package insert that came with the 
Accu-Chek Integra test strip drums.
Measuring range
0.6–33.3 mmol/L (10–600 mg/dL)
Sample size
Approx. 1.5 µL (1 microlitre = 
1 thousandth of a millilitre)
Measuring time
Approx. 15 seconds (depending on 
concentration)
System operating conditions
+10 °C to +40 °C
Meter storage conditions
Without batteries and test strip 
drum: -40 °C to +70 °C
With batteries but without test strip 
drum: -10 °C to +50 °C
With batteries and test strip drum: 
+2 °C to +30 °C
9.2  Troubleshooting