User Manual
9 
Temperature Selection & Monitoring 
Digital Thermostat 
Temperature Parameters: 
MED2 all-refrigerators  come with digital thermostats  that are preset at 4ºC,  with a 
range of 2º to 8ºC. 
VT65MLBIMED2  and  VT65MLBIMED2ADA  all-freezers  are  preset  to  -20º,  with  a 
range of -25º to -10ºC. 
All other MED2 all-freezers are preset to -18ºC, with a range of -20º to -10ºC. 
If you find it necessary to change the temperature setting for your digital thermostat, 
please refer to the following instructions. 
How to Adjust the Temperature: 
To see the SET POINT: 
The default display for your digital thermostat is the internal cabinet temperature. To 
view the unit’s set point, push and immediately release the SET key. The display will 
flash,  showing  the  set  point  value.  If  no  other  actions  are  taken  in  6  seconds,  the 
display will revert to the internal air temperature. 
To change the SET POINT: 
While the set point value is blinking, press the UP or DOWN arrow to change the set 
point to your desired temperature. Press the SET button again to exit the adjustment. 
The  display  will  revert  to  the  internal  air  temperature.  PLEASE  NOTE:  The 
temperature displayed on the digital thermostat is the internal air temperature. Your 
alarm/thermometer (installed on the left in the control panel) provides a temperature 
readout based  on  the buffered probe  in order to better  simulate the  temperature of 
what is actually stored inside the unit (as opposed to the air temperature itself). As a 
result, these readings may not always display the same temperature reading.  
NOTE: 
The minimum and maximum range of your thermostat is preset and 
locked by us, so it cannot easily be accidentally moved outside of the 
recommended temperature ranges. 