ENGLISH Visit www.balances for discount prices on Acculab Scales Introduction Thank you for purchasing an Acculab SVI Series Professional Scale. Please read all of the operating instructions to ensure optimum performance. The unit is pre-calibrated at the factory and is ready to use with the supplied AC adapter. Important Information Allow 15 to 20 minute warm up time before starting calibration routine in order to give internal components sufficient time to stabilize.
ENGLISH Visit www.balances for discount prices on Acculab Scales Key Functions / Description: 5. LIGHT ON / OFF TARE / ZERO F CAL / PRINT 1. 2. 3. 4. 1. ON/OFF: Press to turn scale ON or OFF. 2. ZERO/TARE: Pushing ZERO/TARE key “zero’s” display, pushing and holding ZERO/TARE key for more than 3 seconds accesses application menu. 3. “F” Key: Function key for applications (Unit toggling, Counting & Totaling ) and menu scrolling. 4.
ENGLISH Visit www.balances for discount prices on Acculab Scales 6 Please remove weight and zero scale. 7 Place cal-weight back on to pan and the result should be within +/- 1d. 8 Calibration complete. Calibration Notes: The actual calibration weight needs to be within +/- 5% of the selected weight. If the scale is not zeroed before calibration, error “E-01” will be displayed. If calibration weight is positively “above” the 5% tolerance, error “E-02” will be displayed.
ENGLISH Visit www.balances for discount prices on Acculab Scales 9 To exit “Counting” press and hold “TARE/ZERO” key for more than 3 seconds to access application menu and select “UNIT” (default setting). Totaling: 1 To select the application “Totaling” press and hold “TARE/ZERO” key for more than 3 seconds to access application menu. 2 With the “F” key you can scroll through the menu options (UNIT, COUNT, & TOTAL). 3 Select “TOTAL” by pressing the “CAL/PRINT” Key.
ENGLISH Visit www.balances for discount prices on Acculab Scales Menu Settings: Set up Menu / Configuring Scale 1 2 3 4 5 6 7 8 9 (Factory set parameters are marked “0”.) Turn off scale. Turn on while holding the ZERO/TARE key. Page or level value “1” will be displayed. To navigate within a menu level press “F”, when you reach the last menu item, the first item is shown again. To select the next line or level within a group press and release “CAL/PRINT” key.
ENGLISH Interface Description Type of Interface: Serial Operating Mode: Full Duplex Transmission Rates: 1200, 2400, 4200, 9600 Baud Parity: Mark, Space, Odd, Even Data Output Format: 16 Characters Factory settings: Please See Scale Operating Menu Pin Assignment: Pin2: Tx, Pin3: Rx, Pin5: grnd Commands Print Zero / Tare Change Unit “F” Key “Unit” Mode “Counting” Mode “Totaling” Mode “Calibration” Mode ASCII “p” CR LF “t” CR LF “u” CR LF “f” CR LF “a” CR LF “b” CR LF “c” CR LF “D” CR LF HEX 70 0D 0A 74 0D
7 2 g, 0.002 kg, 0.005 lbs, 0.1 ozt 1 g, 0.001 kg, 0.002 lbs, 0.05 ozt grams, kilograms, pounds, troy ounces To capacity by subtraction 2 kg, 5 kg, 10 kg 200 % “H” reading on LCD with audio tone Displays negative readings Menu selectable: 4 or 10 minutes no activity / Always on 50 - 104 degrees F (10 - 40 degrees C) “LO” on LCD with audio tone Included AC adapter or optional 9 volt battery Stainless steel 308 mm x 308 mm (12.4” x 12.
If you have any questions please contact your local Acculab supplier, visit our website www.acculab.com or contact: Si tiene alguna duda, póngase en contacto con su proveedor local de Acculab, visite nuestra página web www.acculab.com o póngase en contacto con: Wenden Sie sich bei Fragen an Ihren Aculab-Lieferanten vor Ort, besuchen Sie unsere Internetseite www.acculab.com oder wenden Sie sich an: Si vous avez des questions, veuillez contacter votre fournisseur local Acculab, visiter notre site web www.