5.
Alarms
5.1 Alarms and Actions
Description
Display
Reason
Solution
NEAR
alarm
The syringe is
almost empty and
close to end of
infusion
1.press the knob to cancel
the alarm(the alarm will
ring again after two
minutes)
2. Press STOP key to stop
the syringe pump and
cancel the alarm.
OVER
alarm
The syringe is
empty and
infusion is
finished
1. Press the knob to cancel
the alarm(the alarm will
ring again after two
minutes)
2. Press STOP key to stop
the syringe pump and
cancel the alarm.
Occlusion
alarm
1. Syringe or
infusion set
with occlusion
2. Occlusion
alarm need
sensitivity
adjustment
3. Syringe pump
pressure sensor
has problem
1 Press STOP key to
stop the syringe pump and
cancel the alarm, eliminate
infusion occlusion, then
press the START key to
start infusion once again.
No.2 Refer to Occlusion
Sensibility Adjust 3.3
No.3 Contact Digicare
Technical support
Low
battery
1Battery too low
2Bad battery
aging or
something
wrong with
the battery
charge circuit
No.1 Connect to AC
power supply to charge the
battery
No.2 Contact Digicare
Technical Support
AC
indicator is
OFF
1No AC power
supply for the
syringe pump.
2Something
wrong with
the power
supply circuit
of syringe
pump.
No.1Check if the power
cord is not plugged in or
not properly plugged in.
No.2 Contact Digicare
Technical Support