Instruction manual

Neuro Technology, Inc.
Instruction Manual for DigiStim III Peripheral Nerve Stimulator
Part Number: 8-1002-10
TABLE OF CONTENTS
1. INTRODUCTION
2. SETUP
3. FAMILIARIZATION WITH CONTROLS
4. INDICATORS
5. OUTPUT JACKS
6. BATTERY REPLACEMENT
7. APPLICATION NOTES
8. SPECIAL CONSIDERATIONS AND CONTRAINDICATIONS
9. SPECIFICATIONS
1.0 INTRODUCTION
This manual is provided to aid the Anesthesiologists/CRNA in the operation of a Peripheral Nerve Stimulator
(PNS).
1.1 The DIGISTIM III Peripheral Nerve Stimulator is a constant voltage stimulator for monitoring neuromuscular
block and evaluating muscle relaxant dosage during surgery and recovery. The instrument can also be used as an aid
in accurately locating nerves when performing nerve block procedures. A digital display provides an accurate
readout of the current delivered to the patient.
2.0 SETUP
2.1 Location of the instrument should be within six to twelve feet of the patient (depending on leadwire length) and
approximately five feet above the floor to provide easy viewing of the touch switches and indicators. The stimulator
may be placed on an instrument shelf or attached to an IV pole or anesthesia machine using an optional mounting
bracket.
2.2 The leadwires should be attached to the instrument while the numbered control knob is in the OFF position. See
Section 7.1 for instructions on the use of gelled electrodes.
3.0 FAMILIARIZATION WITH CONTROLS
Power is applied to the instrument by rotating the control knob located on the left side of the case in a clockwise
direction until an audible click is heard. When the unit is first turned ON, a random number will appear on the digital
display. The instrument is now in the standby mode and no pulses will be generated. When stimulus pulses are
generated, the digital display will indicate the amplitude of the current in milliamperes (mA). The numbered control
knob is used to adjust the amplitude of the output current. The numbers on the knob serve as a convenient reference
for remembering the approximate settings of the output amplitude.
3.1 The eight membrane touch-switches (buttons) located on the front panel provide a convenient means of
controlling the functions of the instrument. They are operated by using a finger to depress the desired function
designated under each of the eight buttons. Each time a pulse is generated, the Pulse LED will flash and an audible
click will be heard.
3.2 The STANDBY button puts the instrument in standby mode in which no stimulus pulses are generated.
3.3 The TEN SEC button when depressed will generate one pulse every 10 seconds continuously. The 10 second
pulses can be stopped by depressing the STANDBY button or selecting another function.
3.4 The ONE SEC button when depressed will generate one pulse every second continuously. The 1 second pulses
can be stopped by depressing the STANDBY button or selecting another function.
3.5 The TWO HZ button, which is used primarily for nerve location, will cause a pulse to be generated at ½ second
intervals continuously. The 2 Hz pulses can be stopped by depressing the STANDBY button or selecting another
function.
3.6 The 100 HZ TETANUS button is momentary and will deliver 100 pulses per second as long as the button is
depressed.
3.7 The 50 HZ TETANUS button is momentary and will deliver 50 pulses per second as long as the button is
depressed.
3.8 The TOF (Train-Of-Four) button when depressed will automatically generate four pulses in a period of two
seconds. The train-of-four pulses will be repeated every ten seconds until the STANDBY button is depressed or

Summary of content (5 pages)