User`s manual
ProScale M150, M250 and all products with General Purpose LCD Digital Displays  33 of 44 
Pr8 – Automatic Power Off   [0 to 60]  Default = 15.  
Sets the amount of time in ‘minutes without activity’ before the display 
automatically turns off. 
0 = Disables Auto Off.  
Encoder motion or ON/OFF key “wake-ups” the display and resets the timer. 
Pr9 – Auxiliary Keys Enable/Disable  [0..7] 
0 = ABS/INC, MON and HOLD Disabled 
1 = ABS/INC Key Enabled 
2 = MON Key Enabled 
4 = HOLD Key Enabled 
7 = All Keys Enabled 
To enable keys, add up combination of key values. A value of 2 enables only the 
MON key.  A value of 7 enables all 3 Keys. 
Pr10 – Offset Addition Enable  [0, 1] 
0 = Offset Addition Disabled 
1 = Offset Addition Enabled.  
 SEE ALSO Pr11, Pr12, Pr13 
Pr11 – Offset Addition 1 [-999.999 to 999.999in] or [-9999.99 to 9999.99mm]  
When offset 1 is selected (see section 4 -Offset Addition), this value is added to 
the current ABS position. Default = 1.000IN 
Only active if Pr10 is set to 1. Note: Default is set in Inches 
Pr12 – Offset Addition 2 [-999.999 to 999.999in] or [-9999.99 to 9999.99mm]  
When offset 2 is selected (see section 4 -Offset Addition), this value is added to 
the current ABS position. Default: 1.500IN 
Only active if Pr10 is set to 1. Note: Default is set in Inches 
Pr13 – Offset Addition 3 [-999.999 to 999.999in] or [-9999.99 to 9999.99mm]  
When offset 3 is selected (see section 4 -Offset Addition), this value is added to 
the current ABS position and displayed. Default: 2.000IN 
Only active if Pr10 is set to 1. Note: Default is set in Inches 
Pr14 – Output Signal Mode   [0, 1]  
Configures the hardware output signal for activation on MONitor drift conditions or 
Upper/Lower limit alarm conditions. (24VDC displays only) 
0 = Monitor drift, 
1 = Limit.Alarm 
SEE ALSO Pr15, Pr16, Pr17 