Jumpers – Surface Mount Display 
Although the ProScale display uses a keyboard-programming mode to enable and 
configure features in the unit, several selection jumpers are located on the circuit board 
for additional system configuration. 
User configurable jumpers consist of three pins and a ‘shorting block’. 
The center of these pins is ‘Common’. One end pin is labeled ‘A’ and the other end pin is 
labeled ‘B’. 
Surface Mount Display 
Circuit Board 
JP1    FOR FACTORY USE ONLY
JP2    Absolute/Incremental Encoder Selection 
The General Purpose Digital Display supports both Incremental and Absolute style 
measuring systems (See Section 1: ProScale Terminology). To configure the display for 
use with absolute (ABS) type encoders (default), install the shorting jumper in position A.  
For incremental type encoders, install the shorting jumper in position B. 
(ProScale Model 150 & 250 are Absolute systems. Other Accurate Technology products 
such as ProCaliper, ProPanel, Measurement Kits, and ProStop are Incremental systems) 
NOTE: This functionality is not related to the ABS / INC measurement modes described 
in Section 4: Auxiliary Keypad. 
JP3    Programming Enable/Disable 
Entry to the programming mode of the ProScale display can be enabled or disabled 
based on this jumper setting.  To enable keyboard programming (default), install the 
shorting jumper in position A.  To disable keyboard programming, install the shorting 
jumper in position B. When programming mode is disabled, the user cannot access the 
programming functions via the Mode + 0 keys as described in the Section 4: 
Programming. This provides the user with a method of configuring the display with 
specific parameters and prevents unauthorized configuration changes. 
JP4    Display Power 
This jumper configures the Digital Display to operate on either Battery or 24VDC. 
This jumper will be set at the factory based on the type of display you have ordered. 