User`s manual
ProScale M150, M250 and all products with General Purpose LCD Digital Displays  9 of 44 
Incremental Style Pattern 
6.3mm (.25") 
1.9mm (.07”) 
Incremental 
Simply stated, an Incremental system measures the distance it has traveled relative to a 
starting  point.    Incremental style scales  have  a repeating  “bar” pattern on  a  green 
laminate. There should be a “SPLICE” approximately every 570mm (22in.).  Incremental 
scales may be shortened if needed.  
Scales used on Absolute and Incremental systems should never be mixed. 
Avoid  drilling through the green  portion  of any  scale.  Any  portion of  the green  that  is 
drilled will not operate. 
Shortening Scales 
http://www.proscale.com/support/faq.htm 
To shorten incremental scales follow these steps: 
The scale has stainless-steel wires pressed into the side that could be damaged if not cut 
properly. 
1.  Clamp the wires so they do not pull away from the extrusion. 
2.  Using a hacksaw, cut through the aluminum extrusion. 
3.  Use a file or sanding tool to chamfer the cut end to the specifications shown in the 
illustration. 
4.  Remove all burrs. 
5.  Test  the  shortened  scale  by  sliding  a  readhead  on.  There  should  not  be  any 
binding. 
Wires 
Wires 