Installation manual
Manual Part # 800-1402-001 Rev E  Page 31 of 48 
Compensation Functions 
NOTE: The following features are available ONLY on General Purpose Digital 
Readouts with firmware ending in ‘C”, (Part # 700-1600-236) supplied on 
ProScale Model 580 and ProTable systems. 
(The firmware version is displayed on power-up, ie P3.000C). 
TEMPERATURE COMPENSATION  
This feature provides automatic compensation of measurement variations 
caused by changes in the ambient temperature where the measurement 
system is being used. 
Additionally, coefficients of expansion other than aluminum may be 
programmed into the DRO and the resulting measurement compensated for the 
different expansion rates based on a temperature sensor inside the DRO. 
Use this feature to maintain the highest measurement accuracy possible when 
using the system under wide temperature variations or when mounting the 
ProScale aluminum scale extrusion to steel or other surfaces having a different 
temperature expansion rate than aluminum. 
See PROGRAMMING PARAMETERS (Pr41 through Pr44)/ 
To turn on Temperature Compensation: 
1.  Set Pr41 to 1. 
2.  Next, set the compensated Low and High positions using Pr42 & Pr43. 
This step sets the lowest and highest ABS measurement value that will 
be temperature compensated. Most often this will be at 0.000 and the 
maximum measuring range of the system. It is possible however, if the 
system has been placed at a non-zero starting point that the ‘lowest’ 
reading could be a value other than 0.000. 
3.  If necessary, enter the expansion coefficient for the material the 
ProScale aluminum Scale is affixed to in Pr44. 
See Pr44 for details relating to the expansion coefficient setting. 
Use care if changing this setting. It will affect system accuracy! 
Note: Parameters Pr41, 42 & 43 have been preset at the factory for ProTables 
ordered with the “Enhanced Accuracy” option. 